Model Number XS PST |
Device Problem
Erratic or Intermittent Display (1182)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/05/2021 |
Event Type
malfunction
|
Event Description
|
We received an allegation of a display issue with a coaguchek xs meter.The customer reported that the meter only showed three r's after she replaced the batteries.The three r's appeared in the lower section of what would have been the result's field.A display check could not be performed as the meter could not be powered on.But the customer confirmed that all segments were visible prior to the power issue.
|
|
Manufacturer Narrative
|
There were no signs of electrical short circuits.There were no signs of moisture, debris, blood or disinfectant residues in the battery compartment or under the guide cover.It was confirmed that the buttons were not stuck because of contamination.The meter was requested for investigation.Initial reporter occupation: occupation is patient/consumer.
|
|
Manufacturer Narrative
|
The customer's meter was received for investigation.The meter¿s circuit board was tested for damage or contamination and it was found the printed circuit board (pcb) was contaminated.The root cause is determined to be contamination of the contacts due to improper handling or maintenance by the customer.
|
|