Model Number XS PST
Device Problem Erratic or Intermittent Display (1182)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/05/2021
Event Type  malfunction  
Event Description
We received an allegation of a display issue with a coaguchek xs meter.The customer reported that the meter only showed three r's after she replaced the batteries.The three r's appeared in the lower section of what would have been the result's field.A display check could not be performed as the meter could not be powered on.But the customer confirmed that all segments were visible prior to the power issue.
 
Manufacturer Narrative
There were no signs of electrical short circuits.There were no signs of moisture, debris, blood or disinfectant residues in the battery compartment or under the guide cover.It was confirmed that the buttons were not stuck because of contamination.The meter was requested for investigation.Initial reporter occupation: occupation is patient/consumer.
 
Manufacturer Narrative
The customer's meter was received for investigation.The meter¿s circuit board was tested for damage or contamination and it was found the printed circuit board (pcb) was contaminated.The root cause is determined to be contamination of the contacts due to improper handling or maintenance by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key12885604
MDR Text Key285955322
Report Number1823260-2021-03515
Device Sequence Number1
Product Code GJS
UDI-Device Identifier00365702125100
UDI-Public00365702125100
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberXS PST
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/25/2021
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/05/2021
Initial Date FDA Received11/29/2021
Supplement Dates Manufacturer Received11/30/2021
Supplement Dates FDA Received12/21/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-