We checked the returned unit and confirmed that the ccd module defective.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module.In addition, we confirmed that the deflector wire stretched, the us probe failure, the us probe cut, the segment loosened, the light guide fiber bundle (lcb) cut and the us connector cable cut; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.
|