Catalog Number 8065751763
Device Problem Increase in Pressure (1491)
Patient Problems Intraocular Pressure Increased (1937); Capsular Bag Tear (2639)
Event Date 11/03/2021
Event Type  Injury  
Event Description
A nurse reported that during surgery the ophthalmic surgical console presented with high pressure problems and patient experienced with capsule rupture due to pressure peak and patient's eye became hypertonic.Additional information has been requested.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Manufacturer Narrative
The company service representative examined the system and was unable to confirm or replicate the reported event.The system was then tested and met all product specifications.A system manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.The system was found to meet specifications; therefore, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key12891138
MDR Text Key281391730
Report Number2028159-2021-01414
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657517633
UDI-Public00380657517633
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K121555
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number8065751763
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/03/2021
Initial Date FDA Received11/30/2021
Supplement Dates Manufacturer Received01/06/2022
Supplement Dates FDA Received01/25/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/15/2018
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-