Following an atrial fibrillation (af) procedure, a gastroscopy revealed the patient had experienced esophageal necrosis.2d fluoroscopy revealed the necrosis was not localized to just the place where the probe was placed.The temperature cutoff on the probe was unknown.It was noted that the tee probe was removed before the start of the procedure, since the probe was only used for transseptal puncture.The settings during ablation were 30w at the posterior wall and 35w at the anterior wall done by point-by-point ablation with a carto system (biosenes).It was noted that a termocool smart touch* sf ablation catheter was used during the procedure.A few days following the procedure after close tracking via gastroscopy, the patient made a full recovery and is currently in stable condition.Additionally, there were no known patient comorbidities that could have caused or contributed to the reported event.
|