MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. FORCEPS/IRRIGATION PLUG (ISOLATED TYPE); FORCEPS/ IRRIGATION PLUG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. FORCEPS/IRRIGATION PLUG (ISOLATED TYPE); FORCEPS/ IRRIGATION PLUG Back to Search Results
Model Number MAJ-891
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/20/2021
Event Type  malfunction  
Event Description
Olympus medical systems corp.(omsc) confirmed the subject device to investigate the complaint.Omsc found that the brown foreign material adhered to the fitting part of the stopcock with the main body.Other detailed information was not provided.There was no report of patient injury associated with the event.
 
Manufacturer Narrative
The foreign material came from rust and chlorine based on a component analysis.Since the lot number of the subject device was unknown, olympus medical systems corp.(omsc) could not confirm device history record of the subject device.The exact cause of the reported phenomenon could not be conclusively determined.There is the possibility that chlorine come from chemicals that promote oxidation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORCEPS/IRRIGATION PLUG (ISOLATED TYPE)
Type of Device
FORCEPS/ IRRIGATION PLUG
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12906952
MDR Text Key285940340
Report Number8010047-2021-15307
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K912120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberMAJ-891
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/02/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/09/2021
Initial Date FDA Received12/01/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-