Device Problems Break (1069); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to (b)(4).(b)(4) checked the subject device and found the reported phenomenon, and also surmised that this phenomenon was attributed to insufficient cleaning by the user.The exact cause has been under investigation.Therefore, the exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair of the subject device at olympus (b)(4), it was found that there was foreign material on groove or gap of the distal end of the subject device.The occurrence date of the event is unknown, and there was no report of patient injury associated with this event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Olympus medical systems corp.(omsc) has obtained the following additional information from olympus medical systems india (omsi): - the subject foreign material had been found on the forceps elevator.- the ultrasound transducer had a scratch which reached to the adhesive-layer, due to physical stress such as dropping/knocking.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device was returned to olympus medical systems india (omsi).Olympus medical systems corp.(omsc) reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.Omsc could not conclusively determine the exact cause of the foreign material on the forceps elevator since the information from omsi did not contain detailed information about the subject foreign material.In addition, the exact cause of the ultrasound transducer damage could not be conclusively determined.However, based upon the information from omsi, omsc surmised that the subject damage was attributed to the stress, user handling, and/or other factors.If additional information is received, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE
Type of Device
ULTRASOUND GASTROVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12907606
MDR Text Key285941888
Report Number8010047-2021-15316
Device Sequence Number1
Product Code ODG
UDI-Device Identifier04953170341793
UDI-Public04953170341793
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K093395
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberGF-UCT180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/02/2021
Initial Date FDA Received12/01/2021
Supplement Dates Manufacturer Received12/03/2021
01/13/2022
Supplement Dates FDA Received12/27/2021
02/06/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/26/2014
Is the Device Single Use? No