BAXTER HEALTHCARE CORPORATION AK 96; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
|
Back to Search Results |
|
Catalog Number 114364 |
Device Problem
Loose or Intermittent Connection (1371)
|
Patient Problems
Low Blood Pressure/ Hypotension (1914); Nausea (1970); Diaphoresis (2452); Confusion/ Disorientation (2553)
|
Event Date 10/20/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that 1.5 hours after starting treatment with an ak 96 machine, the patient developed sweating, nausea, confusion and blood pressure was 81/60 mmhg.The ultrafiltration was stopped, however symptoms recurred 2 hours after dialysis and the patient was treated with fluid replacement and hemodialysis was stopped.During the patient dialysis, the interface buckle between the water outlet tubing of the substitution solution and the internal connection of the machine got loose, resulting in the possibility of a fluid leak.No alarm condition was reported in connection with the treatment.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the actual device was not available; however, two (2) photographs were provided for evaluation.Based on the event description and on the provided picture, the leak occurred during use from the flow path and in particular, from the connection between the blue dialysis fluid tube to the machine.The cause was identified to be due to part failure.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Search Alerts/Recalls
|
|
|