• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH TRAUMACEM(TM) V+ SYRINGE KIT - STERILE; DISPENSER, CEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH TRAUMACEM(TM) V+ SYRINGE KIT - STERILE; DISPENSER, CEMENT Back to Search Results
Catalog Number 03.702.150S
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/04/2021
Event Type  malfunction  
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021, the patient underwent open reduction internal fixation surgery for trochanteric fracture of femur.In the surgery, the stopcock broke because the surgeon turned a blue syringe attached to the stopcock with excessive force.The surgery was completed with no surgical delay.The patient condition was stable.This complaint involves one (1) device.This report is for (1) traumacem(tm) v+ syringe kit - sterile.This is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Product complaint # (b)(4).Complainant part is not expected to be returned for manufacturer review/ investigation.Initial reporter information unknown.Without a lot number, the device history records review could not be completed as no product was received.The investigation could not be completed, no product was received; no conclusion could be drawn at the time of filing this report.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10: additional narrative: all medical device reports associated with the same/similar device(s) and reported condition of ¿broken" were reviewed.It was determined that ¿broken" has not caused or contributed to any serious injuries or deaths within the reviewed time period of (b)(6) 2019- (b)(6) 2022.Based on the reviewed data, the likelihood of a death or serious injury occurring as a result of ¿broken" is remote; therefore, ¿broken" associated with same/similar device(s) of this report will no longer be reported as a medical device report (mdr) unless the event results in a reportable event per 21 cfr 803.50.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAUMACEM(TM) V+ SYRINGE KIT - STERILE
Type of Device
DISPENSER, CEMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES SELZACH
bohackerweg 5
selzach 2545
SZ   2545
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key12917166
MDR Text Key281605631
Report Number8030965-2021-09894
Device Sequence Number1
Product Code KIH
UDI-Device Identifier07611819476523
UDI-Public(01)07611819476523
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.702.150S
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/03/2021
Initial Date FDA Received12/02/2021
Supplement Dates Manufacturer Received08/15/2022
Supplement Dates FDA Received08/19/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-