Model Number OF-B206
Device Problem Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/15/2021
Event Type  malfunction  
Event Description
Pentax medical was made aware of an event which occurred in the (b)(6) involving pentax accessories suction valves been very sticky.In the event reported, it was stated the suction valves have been very "sticky", and that it is impacting his procedures.He finds that he is having to remove them each time during procedures and dip them in water to have them not be stuck in a "down" position.There was no adverse event reported with this complaint.No other information provided with this complaint.
 
Manufacturer Narrative
This device is classified as import for export, therefore 510k is not applicable.The manufacture reviewed the complaint and notified the user that condition like this complaint case occur when silicon oil is not properly applied to the valves.Dhr cannot be performed since a lot number was not provided.If additional information becomes available, a supplemental report will be filed with the new information.
 
Manufacturer Narrative
Correction information: h6: coding changed, based on the investigation result.G4: premarket identification pma/510(k): model of-b206 is a suction valve, so there is no 510k number.Evaluation summary: although, the cause was investigated, the reason why it could not be used could not be identified.The assumed cause is deterioration of the o-ring, due to silicone oil.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
SUCTION VALVE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller (temporary)
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key12923181
MDR Text Key285988795
Report Number9610877-2021-01706
Device Sequence Number1
Product Code ODG
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
SEE H10
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOF-B206
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/15/2021
Initial Date FDA Received12/03/2021
Supplement Dates Manufacturer Received11/15/2021
Supplement Dates FDA Received12/06/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-