Model Number ALLURA XPER FD10
Device Problem Radiation Output Failure (4027)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/22/2021
Event Type  malfunction  
Event Description
It has been reported to philips that there was a problem with the foot switch and the wifi indicator was red.The system was not in clinical use when the issue was identified.No harm has been reported to philips.Philips has started an investigation of this complaint.
 
Manufacturer Narrative
Philips has investigated this complaint.A philips service engineer inspected the system onsite and identified that there was an intermittent problem with the wireless footswitch.The issue was resolved after re-inserting the battery.When the battery is removed, the power to the footswitch is removed which resets the footswitch.There are two generations of wireless footswitches used in the field.The newest generation has a software bug which can cause loss of wireless connection, and philips initiated medical device correction z-0476-2022/field safety corrective action (2021-igt-bst-020) for this generation.This complaint is related to the previous generation of the wireless footswitch which is not affected by the software bug.Based on the information collected, philips concludes that the problem was likely an electrical issue.Currently, there are no wireless footswitches or base stations available for replacement.The customer has the wired footswitch available as a backup.Updated codes based on investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
222 jacobs st
cambridge, MA 02141
6172455900
MDR Report Key12923361
MDR Text Key281658349
Report Number3003768277-2021-10256
Device Sequence Number1
Product Code IZI
UDI-Device Identifier00884838059030
UDI-Public00884838059030
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K130842
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD10
Device Catalogue Number722010
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/22/2021
Initial Date FDA Received12/03/2021
Supplement Dates Manufacturer Received11/22/2021
Supplement Dates FDA Received03/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/06/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-