DEPUY ORTHOPAEDICS INC US UNKNOWN HIP FEMORAL HEAD Back to Search Results
Catalog Number UNK HIP FEMORAL HEAD
Device Problem Device Dislodged or Dislocated (2923)
Patient Problems Ossification (1428); Arrhythmia (1721); Pain (1994); Joint Dislocation (2374); Joint Laxity (4526)
Event Date 09/06/2016
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided.The information received will be retained for trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Medical records received.Right hip demonstrate cemented right total hip arthroplasty with pressfit cup.After review of medical records patients was revised was due to dislocation, pain and instability and paroxysmal atrial fibrillation.Revision notes reported of scar tissue.Head and poly liner were removed.Cup and stem were well fixed and stable and in good position thus not revised.There is heterotopic ossification.Tripolar construct (bipolar head and poly liner) were implanted.Doi: 2001, dor: (b)(6) 2016, right hip.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned to depuy synthes for evaluation.Review of the x-ray evidence provided found insufficient evidence to confirm an implant issue.No sign of implant dislocation was observed on the evidence provided.The reported condition could not be confirmed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP FEMORAL HEAD
Type of Device
HIP FEMORAL HEAD
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key12923710
MDR Text Key281664530
Report Number1818910-2021-27047
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK HIP FEMORAL HEAD
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/17/2021
Initial Date FDA Received12/03/2021
Supplement Dates Manufacturer Received04/24/2023
Supplement Dates FDA Received04/25/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNK HIP ACETABULAR LINER POLY.; UNKNOWN HIP FEMORAL HEAD.
Patient Outcome(s) Required Intervention;
Patient Age79 YR
Patient SexMale
Patient Weight100 KG
-
-