Model Number 408719
Device Problems Failure to Capture (1081); Ambient Noise Problem (2877)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/24/2021
Event Type  malfunction  
Event Description
Crt interrupt events were reported on home monitor.Pacing configuration is unipolar.Lead values are within normal ranges.During clinic fup on (b)(6) 2021 recreated noise with isometrics.There was no capture in bipolar configurations.X-ray was suggested.Device remains implanted.No adverse patient events were reported.Should additional information become available, this file will be updated.
 
Manufacturer Narrative
The device is currently not available for analysis.No conclusion can be drawn at this time.No additional information is available at the moment.The file is closed.The investigation will be re-opened should additional data become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENTUS PROMRI OTW QP L-85/49
Type of Device
LV LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key12930708
MDR Text Key282294292
Report Number1028232-2021-06527
Device Sequence Number1
Product Code OJX
UDI-Device Identifier04035479148959
UDI-Public04035479148959
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P070008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2022
Device Model Number408719
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/24/2021
Initial Date FDA Received12/04/2021
Supplement Dates Manufacturer Received12/06/2021
Supplement Dates FDA Received12/07/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/25/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age49 YR
Patient SexFemale
-
-