Model Number M00510870 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/12/2021 |
Event Type
Injury
|
Event Description
|
It was reported to boston scientific corporation that a trapezoid rx lithotripter basket was used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2021.During the procedure, an alliance handle was used in conjunction with the trapezoid basket in an attempt to crush a 12mm stone.However, the pull wire broke.The procedure was completed by extracting the stone from the basket using a 15mm balloon.There were no patient complications reported as a result of this event.
|
|
Manufacturer Narrative
|
(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a problem analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
|
|