• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE Back to Search Results
Model Number GF-UCT180
Device Problem Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/21/2021
Event Type  malfunction  
Event Description
As reported customer requested a scope repair with no further details provided.There is no patient involvement on this reported event.No user injury.Device return evaluation found peeling on forceps cover glue.
 
Manufacturer Narrative
The subject device was received and evaluated.Device inspection found forceps cover glue peeling.Furthermore, the following additional findings were noted.A-rubber leaking.A-rubber cut.A-rubber glue (bending section) chipped.S-cover name plate missing.Lg (light guide) tube collapse and buckles.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the cause of the forceps cover glue peeling likely occurred due to an external force applied to the part.Investigation confirmed that there was no adhesive in the position where adhesive should have been applied.The specific root cause could not be determined at this time.The following information is stated in the instructions for use: "inspection of the endoscope 3.Inspect the external surface of the entire insertion section including the bending section and the distal end for dents, bulges, swelling, scratches, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE
Type of Device
ULTRASOUND GASTROVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12942321
MDR Text Key282510653
Report Number8010047-2021-15615
Device Sequence Number1
Product Code ODG
UDI-Device Identifier04953170356339
UDI-Public04953170356339
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K093395
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGF-UCT180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/12/2021
Initial Date FDA Received12/06/2021
Supplement Dates Manufacturer Received12/28/2021
Supplement Dates FDA Received01/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/05/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-