• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THE SPECTRANETICS CORPORATION SPECTRANETICS LEAD LOCKING DEVICE; STYLET, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THE SPECTRANETICS CORPORATION SPECTRANETICS LEAD LOCKING DEVICE; STYLET, CATHETER Back to Search Results
Model Number 518-019
Device Problem Break (1069)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 11/16/2021
Event Type  Injury  
Manufacturer Narrative
Patient's date of birth, age unk.Patient's weight unk.Other relevant history unk.The device was discarded, thus no investigation could be completed.
 
Event Description
A lead extraction procedure commenced to remove a right ventricular (rv) and a right atrial (ra) lead due to non function.Spectranetics lead locking devices (llds) were inserted into each lead to provide traction.The physician began by using a spectranetics 14f glidelight laser sheath to successfully extract the rv lead.The 14f glidelight device was then used to attempt extraction of the ra lead.The glidelight device stalled progress in the superior vena cava (svc) region, and made very slow progress until the patient's blood pressure dropped.Rescue efforts began immediately, including rescue balloon, administration of blood products and medications, and sternotomy.A 1-2 cm perforation was discovered high in the right atrium, but below the svc (mdr #1721279-2021-00234).Patient was placed on bypass and the repair was successfully completed.The physician attempted to unlock the lld from the ra lead, but was unsuccessful.The ra lead tip, containing the lld, was cut and capped and remained in the patient's body.The patient survived the procedure.This report captures the lld which was cut and capped within the ra lead tip and remained in the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS LEAD LOCKING DEVICE
Type of Device
STYLET, CATHETER
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
barbara creel
9965 federal drive
colorado springs, CO 80921
MDR Report Key12957450
MDR Text Key286452902
Report Number1721279-2021-00235
Device Sequence Number1
Product Code DRB
UDI-Device Identifier00813132023010
UDI-Public(01)00813132023010(17)230414(10)FLC21D12A
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K990713
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/14/2023
Device Model Number518-019
Device Catalogue Number518-019
Device Lot NumberFLC21D12A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/16/2021
Initial Date FDA Received12/07/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/12/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
(B)(4) 1944 RA PACING LEAD; (B)(4) 1948 RV ICD LEAD; SPECTRANETICS 14F GLIDELIGHT LASER SHEATH; SPECTRANETICS CVX-300 EXCIMER LASER SYSTEM; SPECTRANETICS LEAD LOCKING DEVICE IN RV LEAD
Patient Outcome(s) Other;
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-