This device is classified as import for export, therefore 510k is not applicable.Model eg38-j10ut-us is available in the usa with a 510k number k200090.We checked the returned unit and confirmed that the ccd driver pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd driver pcb.In addition, we confirmed that the insertion flexible tube (ift) buckled, the light guide fiber bundle (lcb) broken, the deflector operating knob broken, the lg prong corroded, the suction channel buckled, the us connector cable crushed, the lg prong top worn out, and the lg cable connector fluid damage; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.
|