Additional information was added to h3, h4, h6 and h10.H4: device manufacture date: june 10, 2021 - june 11, 2021.H10: the device was received for evaluation.Visual inspection was performed and noted the flow restrictor body was damaged.Leak was confirmed at the damaged flow restrictor body when the folfusor sample was being leak tested with water.The reported condition was verified.The cause of the condition was due to diagonal indentation lines noted on the flow restrictor body which indicated the flow restrictor may have been damaged by the flow wrap sealer equipment during manufacturing.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|