Model Number BI70002000
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Positioning Problem (3009)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/11/2021
Event Type  malfunction  
Manufacturer Narrative
Other relevant device(s) are: product id: bi70002000-g30, serial/lot #: unknown.A manufacturer representative went to the site to test the equipment.It was reported that the dingus was realigned.The imaging system then passed the system checkout and was found to be fully functional.(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that the gantry would not fully close, as it would stay open about one to two inches when holding down the close button on the pendant.  no further details were available.  no patient was present when the issue was identified.On (b)(6) 2021 (rep): no new information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key12963328
MDR Text Key282045951
Report Number3004785967-2021-01357
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00763000132378
UDI-Public00763000132378
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200074
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBI70002000
Device Catalogue NumberBI70002000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/11/2021
Initial Date FDA Received12/08/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/21/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse