MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA Back to Search Results
Model Number 37601
Device Problems Unintended Collision (1429); Low impedance (2285)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2021
Event Type  malfunction  
Event Description
It was reported that electrode 10 & 11 are shorted.This was previous to an implantable neurostimulator (ins) replacement occurring now.The patient did fall a few days ago, but the ins replacement was due to normal battery depletion.The patient is programmed to the 10 electrode currently.The physician may plan on replacing the extension today to see if the short is resolved.No patient symptoms were reported.Additional information was received from the manufacturer representative (rep) reporting that the cause of the short was unknown.The extension was not replaced.The short remains but the device was programmed as a monopolar on 10 with impedances at 850 ohms.This was confirmed with the physician.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key12963969
MDR Text Key288520200
Report Number3004209178-2021-18245
Device Sequence Number1
Product Code MHY
UDI-Device Identifier00763000175719
UDI-Public00763000175719
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/14/2021
Device Model Number37601
Device Catalogue Number37601
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/01/2021
Initial Date FDA Received12/08/2021
Date Device Manufactured07/18/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexMale
Patient Weight93 KG
-
-