Model Number C-2304
Device Problem Fracture (1260)
Patient Problem Insufficient Information (4580)
Event Date 11/11/2021
Event Type  malfunction  
Event Description
The patient was implanted with a lap-band on (b)(6) 2010.The patient had an initial port replacement on (b)(6) 2011.During a ct scan of abdomen and pelvis, a fracture in the port tubing was observed.It was noted that the sleeve on the port came off.The surgeon reported that this may have allowed stress on the port tubing, which broke just after the metal post on the older ez port design.The surgeon stated that the older port sleeve was going too loose to stay on the surrounding coupling, and, therefore, the surgeon was afraid it would happen again and replaced the port on (b)(6) 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAP-BAND SYSTEM
Type of Device
ADJUSTABLE GASTRIC BAND
Manufacturer (Section D)
RESHAPE LIFESCIENCES
1001 calle amanecer
san clemente CA 92673
Manufacturer Contact
michelle ravert
1001 calle amanecer
san clemente, CA 92673
8449377374
MDR Report Key12964931
MDR Text Key285941341
Report Number3013508647-2021-00043
Device Sequence Number1
Product Code LTI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC-2304
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/11/2021
Initial Date FDA Received12/08/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age65 YR
Patient Weight78 KG
-
-