Model Number EG16-K10 |
Device Problems
Partial Blockage (1065); Obstruction of Flow (2423)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/11/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
If additional information becomes available, a supplemental report will be filed with the new information.This device is classified as import for export, therefore 510k is not applicable.
|
|
Event Description
|
(b)(6) 2021 when i was inspecting before use in the morning, there was a blockage in the air supply and water supply.This event occurred at the time of before use.There was no report of patient harm.
|
|
Event Description
|
(b)(6) 2021 when i was inspecting before use in the morning, there was a blockage in the air supply and water supply.This event occurred at the time of before use.There was no report of patient harm.
|
|
Manufacturer Narrative
|
If additional information becomes available, a supplemental report will be filed with the new information.This device is classified as import for export, therefore 510k is not applicable.
|
|
Manufacturer Narrative
|
Evaluation summary: we checked the returned unit and confirmed that the air/water nozzles clogged.Based on the result, we concluded that it was caused due to the insufficient reprocessing at the facility.In addition, we confirmed that the air/water channel clogged, and the lg supply tubes clogged; however, they are not the main cause, and/or irrelevant to the alleged complaint.
|
|