MAUDE Adverse Event Report: COOK ENDOSCOPY ACROBAT¿ CALIBRATED TIP WIRE GUIDE; OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY ACROBAT¿ CALIBRATED TIP WIRE GUIDE; OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Catalog Number ACRO-35-450
Device Problem Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/09/2021
Event Type  malfunction  
Event Description
During an endoscopic retrograde cholangiopancreatography (ercp) procedure, the physician used a cook acrobat¿ calibrated tip wire guide.It was reported that the physician advanced the wire guide to bile duct and found out that the coating of the wire guide peeled off.Physician then changed to another of the same device to complete the procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Manufacturer Narrative
Initial reporter; occupation: unknown.The product was received for evaluation and the investigation is on-going.A follow-up emdr report will be provided.
 
Event Description
During an endoscopic retrograde cholangiopancreatography (ercp) procedure, the physician used a cook acrobat¿ calibrated tip wire guide.It was reported that the physician advanced the wire guide to bile duct and found out that the coating of the wire guide peeled off.Physician then changed to another of the same device to complete the procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Manufacturer Narrative
Initial reporter; occupation: unknown.The product was received for evaluation and the investigation is on-going.A follow-up emdr report will be provided.
 
Manufacturer Narrative
Section e.Initial reporter; occupation: unknown.Investigation evaluation: our evaluation of the product said to be involved confirmed the report.There is wire guide coating damage near the distal end.A section of core wire was exposed approximately 25.9cm to 27.7cm from the distal end.From approximately 25.3cm to 25.9cm area the coating has folded over itself multiple times.Due to the condition of the returned device it cannot be determined if any sections of the coating are missing.A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis.¿ the device history record for the lot number said to be involved was reviewed.The device history record contains nonconformances that could potentially be related to coating damage.The device goes through various inspections prior to leaving the facility.These inspections would have removed any nonconforming devices prior to distribution.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.If additional pressure is applied to the wire guide and/or accessory device(s) while moving the wire guide inside the accessory device(s), this could contribute to wire guide coating damage.Prior to distribution, all acrobat¿ calibrated tip wire guides are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACROBAT¿ CALIBRATED TIP WIRE GUIDE
Type of Device
OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key12967579
MDR Text Key286351188
Report Number1037905-2021-00673
Device Sequence Number1
Product Code OCY
UDI-Device Identifier10827002342665
UDI-Public(01)10827002342665(17)230616(10)W4355472
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/16/2023
Device Catalogue NumberACRO-35-450
Device Lot NumberW4355472
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2021
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/11/2021
Initial Date FDA Received12/09/2021
Supplement Dates Manufacturer Received12/09/2021
Supplement Dates FDA Received01/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/16/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
OLYMPUS JF-260V ENDOSCOPE
Patient Age12 YR
Patient SexFemale
Patient Weight40 KG
-
-