Model Number 777F8
Device Problem Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/14/2021
Event Type  malfunction  
Event Description
It was reported that before use, the svo2 calibration could not be performed.It was unknown what type of monitor was used and if any error message was displayed on the monitor.There was no injury or complications to the patient.During the product evaluation, it was revealed that the balloon did not inflate due to leakage from a tear.The torn edges were not able to match up.
 
Manufacturer Narrative
One 777f8 catheter without any attached components was returned for examination.The reported issue of the "svo2 calibration could not be performed" was unable to be confirmed.The catheter passed the in-vitro calibration on a vigilance ii monitor.All through lumens were patent without any leakage or occlusion.The balloon did not inflate due to leakage from a tear, approximately 1.5 mm x 0.5 mm in size, along the distal balloon bonding site.The torn edges were not able to match up.There was no visible damage observed from catheter body.The sample was sent for further evaluation.The catheter passed the attenuation test.A review of the manufacturing records indicated that the product met specifications upon release.As part of the manufacturing process, 100% balloon inflation and visual inspection is performed.A product risk assessment was initiated to address the failure complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ CCOMBO V CCO/SVO2/CEDV/VIP THERMODILUTION CATHETER
Type of Device
CATHETER, OXIMETER, FIBER-OPTIC
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
jessica atallah
1 edwards way
irvine, CA 92614
9492500294
MDR Report Key12972440
MDR Text Key283236550
Report Number2015691-2021-06750
Device Sequence Number1
Product Code DQE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K040287
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/14/2022
Device Model Number777F8
Device Catalogue Number777F8
Device Lot Number63434323
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/19/2021
Initial Date FDA Received12/09/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/17/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-