HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO ADAPTER MODULE EPK W/OUT EYEPC Back to Search Results
Model Number PVK-1070Z
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/19/2021
Event Type  malfunction  
Manufacturer Narrative
This device is not distributed in us so that 510k# is blank.We checked the returned unit and confirmed the ccd module as defective.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.
 
Event Description
The time of event is unknown.There was no report of patient harm.Video image failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO ADAPTER MODULE EPK W/OUT EYEPC
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
MDR Report Key12974075
MDR Text Key286531730
Report Number9610877-2021-50742
Device Sequence Number1
Product Code FEA
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 12/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPVK-1070Z
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/19/2021
Initial Date FDA Received12/10/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-