Envoy medical corp.(emc) was notified on 10/28/2021 of a patient reporting eii device cutting in and out immediately after a battery change procedure (occurring (b)(6) 2021).A follow-up procedure was completed (b)(6) 2021 and a backed-out driver lead was confirmed.The sp was removed and replaced with a new sp.The system was then found to be operating as expected after the revision procedure.Patient is meeting nal-rp target with excellent wrs.No device or system performance issues are indicated or alleged.Patient/clinical history with emc: (b)(6) 2012 ,105-0123(a0) st-mo implant; (b)(6) 2012, 105-0123(a0) st-mo activation; (b)(6) 2012, 105-0123(a0) st-mo fitting; (b)(6) 2017, 105-0123 stmo fitting; (b)(6) 2017, 105-0123 st-mo battery change; (b)(6) 2021, s1051-105-0123 st-mo ibc (driver lead backed out); (b)(6) 2021, 105-0123 st0mo post ibc fitting; (b)(6) 2021, s1051 105-0123 st-mo ibc to address driver lead backing out; (b)(6) 2021, 105-0123 st-mo post ibc fitting.
|