Other, other text: this remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No problem or issues were identified during the device history record review.Samples were returned for evaluation.The samples were visually inspected under normal conditions of illumination according to site visual inspection procedure.They then were aligned to perform quality check as procedure, there was no variation appreciable in none of the sample groups.The length of returned samples were checked using the length gauge as procedure.One of the three decontaminated samples passed, while the other two were rejected; no discrepancies were detected on unused samples, both passed the length check.The most probable root cause is a human error occurred after cutting the cannula to the required length while dispositioning to scrap, since the calibrated measuring fixture gauge can detect a dimensional issue.
|