DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problems
Erratic or Intermittent Display (1182); Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/17/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate the iabp and was unable to reproduce the reported issue.All functional and safety checks were performed to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported during an in house training, the cardiosave intra-aortic balloon pump (iabp) display was blinking and freezing and the buttons were not functioning.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Updated fields; b4, g3, g6, g7, h2, h4, h6, h10, h11.Corrected fields: g1.
|
|
Event Description
|
N/a.
|
|