Model Number 96102
Device Problem No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/24/2021
Event Type  malfunction  
Manufacturer Narrative
Spacelabs has launched an investigation and will file a supplemental report when the investigation is complete.
 
Event Description
Spacelabs received a report from the monitor tech that on (b)(6) 2021, the telemetry system went blank, and it did not came back.No injury was reported by the hospital for this event.
 
Manufacturer Narrative
The monitor was received at spacelabs¿ equipment service center for evaluation.The hard drive and power supply replaced resolving the reported issue.Monitor returned to customer site.This report is complete, and this particular issue is considered closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPACELABS XHIBIT CENTRAL STATION
Type of Device
XHIBIT CENTRAL MONITOR
Manufacturer (Section D)
SPACELABS HEALTHCARE INC.
35301 se center st.
snoqualmie WA 98065
Manufacturer Contact
marc rivas
35301 se center st.
snoqualmie, WA 98065
4253635568
MDR Report Key12981123
MDR Text Key284815685
Report Number3010157426-2021-00070
Device Sequence Number1
Product Code MHX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122146
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/01/2005,02/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number96102
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Initial Date Manufacturer Received 11/24/2021
Initial Date FDA Received12/10/2021
Supplement Dates Manufacturer Received11/24/2021
Supplement Dates FDA Received02/04/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-