Model Number 96102 |
Device Problem
No Display/Image (1183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/24/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Spacelabs has launched an investigation and will file a supplemental report when the investigation is complete.
|
|
Event Description
|
Spacelabs received a report from the monitor tech that on (b)(6) 2021, the telemetry system went blank, and it did not came back.No injury was reported by the hospital for this event.
|
|
Manufacturer Narrative
|
The monitor was received at spacelabs¿ equipment service center for evaluation.The hard drive and power supply replaced resolving the reported issue.Monitor returned to customer site.This report is complete, and this particular issue is considered closed.
|
|