• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROMODULATION ACTIVA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROMODULATION ACTIVA Back to Search Results
Model Number 37612
Device Problems High impedance (1291); Unintended Collision (1429)
Patient Problems Muscular Rigidity (1968); Shaking/Tremors (2515)
Event Date 12/09/2021
Event Type  malfunction  
Event Description
It was reported that the patient had a return of symptoms on the right side of his body.The patient experienced a fall 2 days ago.An impedance test was conducted and high impedances were found on all contacts except the 0 in mono-polar configuration.The healthcare provider (hcp) ordered x-ray and ct to determine if there was a lead fracture.Patient was reprogrammed using c+, 0- and the patient had improved control of tremor and rigidity.The issue was not resolved at this time.
 
Manufacturer Narrative
Concomitant medical products: product id: 3389s-40, lot#: va0cnv2, implanted: (b)(6) 2013, product type: lead; product id: 3708660, serial# (b)(4), implanted: (b)(6) 2013, product type: extension.Other relevant device(s) are: product id: 3389s-40, serial/lot #: (b)(4), ubd: 07-aug-2016, udi#: (b)(4); product id: 3708660, serial/lot #: (b)(4), ubd: 11-jul-2017, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from a manufacturer representative (rep) reporting that the doctor reviewed the x-ray and ct and did not see evidence of a lead fracture.The high impedances did not resolve but the patient's symptoms were resolved after reprogramming around the electrodes involved in the impedances.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
7000 central ave ne
minneapolis MN 55432
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central ave ne
minneapolis MN 55432
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key12984700
MDR Text Key285490161
Report Number2182207-2021-02156
Device Sequence Number1
Product Code MHY
UDI-Device Identifier00643169529762
UDI-Public00643169529762
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/14/2016
Device Model Number37612
Device Catalogue Number37612
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/09/2021
Initial Date FDA Received12/13/2021
Supplement Dates Manufacturer Received12/13/2021
Supplement Dates FDA Received12/20/2021
Date Device Manufactured01/21/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age74 YR
Patient SexMale
-
-