• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH CABLE CUTTER-STANDARD; CUTTER, WIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH CABLE CUTTER-STANDARD; CUTTER, WIRE Back to Search Results
Catalog Number 391.905
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Initial reporter is a j&j employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that the cable cutter was discovered broken during a surgery and therefore was not used in the procedure.No further information provided.This report is for one (1) cable cutter-standard.This is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: part: 391.905 lot: 82p3670 manufacturing site: tuttlingen release to warehouse date: march 10, 2021 a review of the device history records was performed for the finished device lot number, and no non-conformances were identified.The raw material certificate was reviewed and the used material was according to the specification of the device.Visual inspection: the cable cutter std was received at us customer quality (cq).Visual inspection of the complaint device showed that one of the jaw of cable cutter was broken and broken fragments were not returned.No other defects were found with the device.Dimensional inspection: no dimensional inspection was performed due to the complaint condition.Document/specification review: current & manufactured revisions were reviewed.No design issues or discrepancies were identified.Investigation conclusion: this complaint is confirmed as the one of the jaw was broken.No root cause could definitively be determined for the reported complaint condition.No new, unique, or different patient harms were identified as a result of this evaluation.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventive action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CABLE CUTTER-STANDARD
Type of Device
CUTTER, WIRE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES TUTTLINGEN
unter hasslen 5
tuttlingen 78532
GM   78532
Manufacturer Contact
kara ditty-bovard
1302 wrights lane east
west chester, PA 19380
6103142063
MDR Report Key12984831
MDR Text Key286642552
Report Number8030965-2021-10161
Device Sequence Number1
Product Code HXZ
UDI-Device Identifier07611819719873
UDI-Public(01)07611819719873
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number391.905
Device Lot Number82P3670
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/30/2021
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/17/2021
Initial Date FDA Received12/13/2021
Supplement Dates Manufacturer Received12/30/2021
01/26/2022
Supplement Dates FDA Received01/25/2022
02/15/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/10/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-