ARROW INTERNATIONAL INC. ARROW HEMODIALYSIS SET: 2-LUMEN 12 FR X 20 CM; CATHETER HEMODIALYTSIS NON IMP Back to Search Results
Catalog Number CS-25122-F
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/20/2021
Event Type  malfunction  
Event Description
It was reported the extension line was found cracked during patient use.No patient harm reported.The patient's condition is reported as fine.
 
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported the extension line was found cracked during patient use.No patient harm reported.The patient's condition is reported as fine.
 
Manufacturer Narrative
(b)(4).The customer returned one 3-l cvc catheter for analysis.Signs of use in the form of biological material was observed on the extension lines.The separated portion of the distal extension line was not returned.Visual inspection confirmed the distal luer hub separated from the distal extension line.The separation point was rough and jagged.The customer report of an extension line separation could be confirmed through visual inspection of the returned sample.The distal luer hub separated from its extension line.However, without the luer hub returned for analysis, the point of separation could not be determined.Based on the customer report, the incorrect material number, and the missing luer hub, the root cause could not be determined.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW HEMODIALYSIS SET: 2-LUMEN 12 FR X 20 CM
Type of Device
CATHETER HEMODIALYTSIS NON IMP
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key12992598
MDR Text Key282284503
Report Number3006425876-2021-01147
Device Sequence Number1
Product Code MPB
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K993933
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date08/20/2022
Device Catalogue NumberCS-25122-F
Device Lot Number71F20H2873
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/24/2021
Initial Date FDA Received12/14/2021
Supplement Dates Manufacturer Received03/18/2022
Supplement Dates FDA Received03/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/14/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NOT REPORTED; NOT REPORTED
-
-