Catalog Number UNKNOWN
Device Problems Loss of or Failure to Bond (1068); Disconnection (1171)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/22/2021
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event Description
It was reported that the patch in the foley catheter that was holding the tubing to the leg bag fell off.Per additional information received via liberator on 01dec2021, it appeared that the patient's allegation was actually against the stabilization device.
 
Event Description
It was reported that the patch in the foley catheter that was holding the tubing to the leg bag fell off.Per additional information received via liberator on 01dec2021, it appeared that the patient's allegation was actually against the stabilization device.
 
Manufacturer Narrative
The reported event was inconclusive because no sample was returned for evaluation.A potential root cause for this failure could be due to ¿adhesive chemistry not appropriate for application (poor quick stick characteristics)¿.It was unknown whether the device had met specifications.The product was used for treatment purpose.It was unknown whether the product had caused the reported failure.The dhr review could not be performed without a lot number.Based on the results of the investigation, no additional actions are needed.The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.Although the product family is unknown, the statlock ifus are found to be adequate based on past reviews.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STATLOCK SL FOLEY, UNKNOWN CODE
Type of Device
UNKNOWN STATLOCK DEVICE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key12994702
MDR Text Key284816759
Report Number1018233-2021-08019
Device Sequence Number1
Product Code EYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/01/2021
Initial Date FDA Received12/14/2021
Supplement Dates Manufacturer Received03/18/2022
Supplement Dates FDA Received04/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-