• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MO016 MO-MOBERLY HOT PACK, LARGE, 6X9IN; PACK, HOT OR COLD, DISPOSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MO016 MO-MOBERLY HOT PACK, LARGE, 6X9IN; PACK, HOT OR COLD, DISPOSABLE Back to Search Results
Model Number 11443-012B
Device Problem Burst Container or Vessel (1074)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/17/2021
Event Type  malfunction  
Manufacturer Narrative
At the time of this investigation the device history record could not be verified, as a lot number was not provided.As no samples were returned for evaluation, a root cause could not be determined.However, a warning label is being added to the label that specifies, ¿when activating, hold away from patient.¿ we will continue to monitor complaint trends and utilize the information as part of continuous improvement.
 
Event Description
Customer reported cardinal health instant hot pack exploded on activation on the gloves of the nurse.No further information was provided when requested.Report being filed for risk.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOT PACK, LARGE, 6X9IN
Type of Device
PACK, HOT OR COLD, DISPOSABLE
Manufacturer (Section D)
MO016 MO-MOBERLY
808 w highway 24
moberly MO 65270
Manufacturer (Section G)
MO016 MO-MOBERLY
808 w highway 24
moberly MO 65270
Manufacturer Contact
patricia tucker
1500 waukegan rd
waukegan, IL 60085
8478874151
MDR Report Key13003407
MDR Text Key284821757
Report Number1423537-2021-00698
Device Sequence Number1
Product Code IMD
UDI-Device Identifier10885380169267
UDI-Public10885380169267
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial
Report Date 12/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number11443-012B
Device Catalogue Number11443-012B
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/17/2021
Initial Date FDA Received12/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
-
-