The device was evaluated by olympus.The evaluation confirmed the reported event of device not flexing up or down.The evaluation found a brownish liquid coming out from bending section sheath and the device image was very cloudy.Additionally, the device scope connector was corroded due to fluid invasion and all switches on the device are not working.The faulty parts were replaced to meet olympus functional standard.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
|
A user facility returned the olympus, uretero- reno videoscope due to no angulation and control lever not flexing up or down.Upon inspection and testing of the returned device, it was observed that the bending manipulation and insertion tube were defective.Additionally, a brown liquid was coming out from the bending section cover.The bending rubber was broken, torn and a metal was sticking out of the device.There was no harm or user injury reported due to the event.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and correction to g2.G2 - checked health professional.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the cause of the events likely occurred from user applying excessive force to the device.The specific root cause could not be determined at this time.The following information is stated in the instructions for use regarding handling the device which may have prevented the event: "precautions : perform a leakage test on the endoscope after each precleaning procedure.Do not use the endoscope if a leak is detected.Use of an endoscope with a leak may cause a sudden loss of the endoscopic image, damage to the bending mechanism, or other malfunctions.Use of a leaking endoscope may also pose an infection control risk.Inspection of the endoscope : inspect the external surface of the entire insertion section, including the bending section and the distal end for dents, bulges, swelling, scratches, peeling of coating, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities.If the movement of the up/down angulation lock and the angulation control lever is loose and/or not smooth, or the bending section does not angulate smoothly, the bending mechanism may be abnormal.In this case, do not use the endoscope because it may be impossible to straighten the bending section during an examination, and patient injury, bleeding, and/or perforation may result.Do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist.Do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damaged.Important information¿please read before use - warnings and cautions - _ do not strike, bend, hit, pull, twist, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, or endoscope connector of the endoscope with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding and/or perforations.It could also cause parts of the endoscope to fall off inside the patient." olympus will continue to monitor field performance for this device.
|