MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V3R
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/10/2021
Event Type  malfunction  
Manufacturer Narrative
The device was evaluated by olympus.The evaluation confirmed the reported event of device not flexing up or down.The evaluation found a brownish liquid coming out from bending section sheath and the device image was very cloudy.Additionally, the device scope connector was corroded due to fluid invasion and all switches on the device are not working.The faulty parts were replaced to meet olympus functional standard.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Event Description
A user facility returned the olympus, uretero- reno videoscope due to no angulation and control lever not flexing up or down.Upon inspection and testing of the returned device, it was observed that the bending manipulation and insertion tube were defective.Additionally, a brown liquid was coming out from the bending section cover.The bending rubber was broken, torn and a metal was sticking out of the device.There was no harm or user injury reported due to the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and correction to g2.G2 - checked health professional.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the cause of the events likely occurred from user applying excessive force to the device.The specific root cause could not be determined at this time.The following information is stated in the instructions for use regarding handling the device which may have prevented the event: "precautions : perform a leakage test on the endoscope after each precleaning procedure.Do not use the endoscope if a leak is detected.Use of an endoscope with a leak may cause a sudden loss of the endoscopic image, damage to the bending mechanism, or other malfunctions.Use of a leaking endoscope may also pose an infection control risk.Inspection of the endoscope : inspect the external surface of the entire insertion section, including the bending section and the distal end for dents, bulges, swelling, scratches, peeling of coating, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities.If the movement of the up/down angulation lock and the angulation control lever is loose and/or not smooth, or the bending section does not angulate smoothly, the bending mechanism may be abnormal.In this case, do not use the endoscope because it may be impossible to straighten the bending section during an examination, and patient injury, bleeding, and/or perforation may result.Do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist.Do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damaged.Important information¿please read before use - warnings and cautions - _ do not strike, bend, hit, pull, twist, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, or endoscope connector of the endoscope with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding and/or perforations.It could also cause parts of the endoscope to fall off inside the patient." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13010687
MDR Text Key282290881
Report Number8010047-2021-16177
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170403392
UDI-Public04953170403392
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V3R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/18/2021
Initial Date FDA Received12/15/2021
Supplement Dates Manufacturer Received01/17/2022
Supplement Dates FDA Received01/27/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/08/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-