This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and correction to g2.G2 - checked "other" to add the country belgium.Three attempts were performed to obtain the cleaning, disinfection, and sterilization of the scope instructions but were not successful.If additional information is obtained at a later date, a supplemental report will be submitted.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is unlikely the positive culture was caused by the device.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with instructions for use (ifu), the results conformed to the regulation's recommendation.This information is addressed in the instructions for use (ifu): "warning: thoroughly clean and high-level disinfect or sterilize the endoscope before returning it for repair.Improperly reprocessed equipment poses an infection control risk to each person who handles the endoscope within the hospital or at olympus." olympus will continue to monitor the field performance of this device.
|