Model Number 3100A
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/18/2021
Event Type  malfunction  
Manufacturer Narrative
Vyaire medical file identification: pc-(b)(4).The customer troubleshoots the unit took same circuit and used on another unit and it passed fine.The customer bypassed humidifier.Technical support suggest to check to see where the limit is set at and look at needle valves, unit has 5165 hours.For worst case customer need to replaced pr3 and technical support provide pn and price.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer reported to vyaire medical that the 3100a ventilator cannot get map (mean airway pressure) higher than 18cm with pr3 maxed out.The customer confirmed that there is no patient involvement associated with this reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
510 technology dr
irvine IL 92618
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
erika bonilla
510 technology dr
irvine, IL 92618
7149227837
MDR Report Key13015048
MDR Text Key282610903
Report Number2021710-2021-15161
Device Sequence Number1
Product Code LSZ
UDI-Device Identifier10846446003079
UDI-Public(01)10846446003079(11)20141001
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/18/2021
Initial Date FDA Received12/15/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown