BAXTER HEALTHCARE CORPORATION INTERLINK BLOOD RECIPIENT SET; SET, BLOOD TRANSFUSION Back to Search Results
Catalog Number 2C6750H
Device Problem Filling Problem (1233)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/22/2021
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the soft filter of four (4) interlink system y-type blood/solution sets was clogging and slowed the flow.The issue was identified during patient infusion.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information was added to h3, h4, h6, and h10.H10: the actual device was not available; however, a companion sample was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing including clear passage and pressure tests were performed and the device operated according to product specifications.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLINK BLOOD RECIPIENT SET
Type of Device
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
carretera sanchez km 18.5
parque industrial itabo, piisa
haina, san cristobal 91000
DR   91000
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key13015589
MDR Text Key282830393
Report Number1416980-2021-07376
Device Sequence Number1
Product Code BRZ
UDI-Device Identifier00085412073583
UDI-Public(01)00085412073583
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K993120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C6750H
Device Lot NumberDR21F25086
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/18/2021
Initial Date FDA Received12/16/2021
Supplement Dates Manufacturer Received12/21/2021
Supplement Dates FDA Received01/10/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/28/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NI.
Patient SexFemale
-
-