PHILIPS MEDICAL SYSTEMS DMC GMBH DIGITALDIAGNOST 4 HIGH PERFORMANCE; SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER) Back to Search Results
Model Number 712027
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer complained that the brake cylinder broke out of the frame of the patient support for stitching.The brake cylinder prevents the footboard of the patient support for stitching from falling down unexpected.A falling footboard can lead in worst case to a fracture in the foot.No injury occurred.
 
Manufacturer Narrative
Ref.Id: (b)(4).The investigation is still ongoing on this event.When the investigation is completed a follow-up report will be sent to the fda.
 
Manufacturer Narrative
Ref.Id: (b)(6).The digitaldiagnost is a direct digital radiography system with flat detector technology based on modular components to allow customization for all radiographic applications and workload requirements for anatomies that are larger than the detector size like scoliosis, it is possible to make a series of exposures covering the whole anatomy (for example full spine or full legs).These individual images can be "stitched" together via the software program.For proper patient positioning and support, the patient can be placed on the so called patient support for stitching or stitching stand, a platform with two handles that the patient can hold on to during the examination.For the ease of use during transportation, the patient support for stitching has wheels and a folding footboard.The footboard is connected to the frame via two hinges and a brake cylinder.It has to be folded up and fixed by a hook for transportation, e.G.From one room to another.Philips field service engineer visited the customer site.The mounting of brake cylinder was not broken.The brake cylinder has become detached from the stitching stand.The cause of this issue is most probably wear and tear.Philips field service engineer replaced the whole patient support for stitching for a newer version.System works as specified.Risk estimation revealed no unacceptable risk.The issue is further monitored and trended.The event was originally reported to the authorities as not enough information was available to make a definite reportability decision.Since that time, additional information was received which revealed that the event does not meet the criteria for reporting.Correction: h6 result and conclusion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIGITALDIAGNOST 4 HIGH PERFORMANCE
Type of Device
SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS DMC GMBH
roentgenstrasse 24
hamburg 22335
GM  22335
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS DMC GMBH
roentgenstrasse 24
hamburg 22335
GM   22335
Manufacturer Contact
dusty leppert
222 jacobs st
cambridge, MA 02141
6172455900
MDR Report Key13018355
MDR Text Key282342996
Report Number3003768251-2021-10018
Device Sequence Number1
Product Code MQB
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K141736
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number712027
Device Catalogue Number712027
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/07/2021
Initial Date FDA Received12/16/2021
Supplement Dates Manufacturer Received12/07/2021
Supplement Dates FDA Received04/07/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-