• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES MANSFIELD CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES MANSFIELD CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS Back to Search Results
Catalog Number 261221
Device Problem Fail-Safe Did Not Operate (4046)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2021
Event Type  malfunction  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported that during the use of a perforator, the automated stop mechanism failed and continued drilling through the bone.No patient harm.
 
Event Description
N/a.
 
Manufacturer Narrative
Additional information received: - date of event :(b)(6) 2021.- due to surgical caution and support of drill the dura was breached but no damage to pia or the cerebral cortex.- the procedure was completed.
 
Manufacturer Narrative
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h10.The perforator was returned for evaluation: dhr - there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Failure analysis - the perforator unit was inspected using the unaided eye.Unit was lightly soiled but showed no other anomalies.Ifu testing procedure was performed after freeing the hudson end which showed slight resistance.Once freed it performed as intended.Functional testing was performed using the same protocol it underwent at finished goods testing prior to release: the unit was found to perform as intended and fulfilled the acceptance criteria.The complaint could not be verified through failure analysis.The root cause is undetermined and was unable to be confirmed in the complaint evaluation.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key13022724
MDR Text Key289529254
Report Number3014334038-2021-00261
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number261221
Device Lot Number5869353
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/17/2022
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/07/2021
Initial Date FDA Received12/16/2021
Supplement Dates Manufacturer Received01/10/2022
02/18/2022
Supplement Dates FDA Received01/26/2022
02/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/21/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-