We checked the returned unit and confirmed the ccd module as defective.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module.In addition, we confirmed that the operation channel leakage, the air/water nozzles loosened, the adjusting collar ruptured, the objective gluing missing, the control body corroded, the remote control buttons cut, and the r/l lock knob improper adjustment; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.
|