STRYKER CORPORATION STRYKEFLOW; APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED
|
Back to Search Results |
|
Model Number 0250070520 |
Device Problems
Fluid/Blood Leak (1250); Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/05/2021 |
Event Type
malfunction
|
Event Description
|
The tubing of the suction irrigator from the strykeflow2 pump disconnected (where it is not supposed to disconnect) dripping fluid, also opening a closed sterile system possibly introducing "non-sterile" pathogens into the patient's abdomen during irrigation.The dripping was noticed within 2-5 minutes after attaching the bag of fluid and immediate discontinued use and replaced.Defective item and packaging sent to qa for the department.
|
|