Model Number 82406 |
Device Problems
High Readings (2459); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/21/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation: the run data file (rdf) was analyzed for this event.The signals in the run data file indicate that the higher-than-expected wbc content in the platelet product was likely a result of an escape of wbcs from the lrs chamber that occurred toward the latter part of the collection.Based on the signals, it cannot be ruled out that this disruption may have been due to an occlusion or air block in the platelet line in the centrifuge that disrupted the steady state of the system.Additionally, it also cannot be ruled out that the higher-than-expected wbc content in the platelet product could be donor-related.The device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.Investigation is in process, a follow-up report will be provided.
|
|
Event Description
|
The customer would like the following run data file investigated to determine a possible cause for the higher-than-expected white blood cell (wbc) content in the platelet product.Donor unit id: (b)(6) there was not a transfusion recipient or patient involved at the time of the residual wbc testing, therefore no patient information is reasonably known at the time of the event.The platelet collection set is not available for return because it was discarded by the customer.
|
|
Event Description
|
The customer would like the following run data file investigated to determine a possible cause for the higher-than-expected white blood cell (wbc) content in the platelet product.Donor unit id: (b)(6) there was not a transfusion recipient or patient involved at the time of the residual wbc testing, therefore no patient information is reasonably known at the time of the event.The platelet collection set is not available for return because it was discarded by the customer.
|
|
Manufacturer Narrative
|
This report is being filed to provide additional information in h.6 and h.10.Investigation: the run data file (rdf) was analyzed for this event.The signals in the run data file indicate that the higher-than-expected wbc content in the platelet product was likely a result of an escape of wbcs from the lrs chamber that occurred toward the latter part of the collection.Based on the signals, it cannot be ruled out that this disruption may have been due to an occlusion or air block in the platelet line in the centrifuge that disrupted the steady state of the system.Additionally, it also cannot be ruled out that the higher-than-expected wbc content in the platelet product could be donor-related.The device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.Root cause: the signals in the run data file indicate that the higher-than-expected wbc content in the platelet product was likely a result of an escape of wbcs from the lrs chamber that occurred toward the latter part of the collection.Based on the signals, it cannot be ruled out that this disruption may have been due to an occlusion or air block in the platelet line in the centrifuge that disrupted the steady state of the system.Additionally, it also cannot be ruled out that the higher-than-expected wbc content in the platelet product could be donor-related.
|
|
Search Alerts/Recalls
|