• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION MEDTRONIC HALL PROSTHETIC HEART VALVE; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION MEDTRONIC HALL PROSTHETIC HEART VALVE; HEART-VALVE, MECHANICAL Back to Search Results
Model Number M7700
Device Problem Mechanical Jam (2983)
Patient Problems Non specific EKG/ECG Changes (1817); Low Blood Pressure/ Hypotension (1914)
Event Date 02/01/1988
Event Type  Death  
Manufacturer Narrative
Citation: masters rg, et al.Extrinsic obstruction of the medtronic-hall disk valve in the mitral position.Ann thorac surg.1988 fe b;45(2):210-2.Doi: 10.1016/s0003-4975(10)62440-x.Earliest date of publish used for date of event and date of death: february 1, 1988 (month and year valid).No unique device identifier (serial/lot) numbers were provided; without this information it could not be determined whether these observations have been previously reported.Without return of the product no definitive conclusion can be made regarding the clinical observations.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information from a literature article regarding three cases of extrinsic obstruction of the medtronic-hall prosthesis in the mitral position.Patient 1: a (b)(6) female patient who underwent mitral valve replacement with a 29 mm medtronic-hall mechanical valve using an interrupted mattress suture technique of ethibond buttressed with teflon pledgets.The valve seated well, with free movement of the disk occluder.The patient was weaned from cardiopulmonary bypass without difficulty.However, during closure of the sternotomy, hypotension occurred with electromechanical dissociation, necessitating reinstitution of cardiopulmonary bypass.The left atrium was re-opened, and the disk occluder was found to be firmly stuck in the closed position.One suture had torn through the friable anterolateral mitral annulus and was overhanging and had become lodged between the valve ring and disk, jamming the valve in the closed position.The disk was released manually, and the valve was re-sutured with considerable difficulty.Despite intra-aortic balloon counter-pulsation and dopamine therapy, the patient could not be weaned from cardiopulmonary bypass.The authors did not explicitly state the patient¿s outcome, but based on the given circumstances, it is likely the patient died.Patient 2: a (b)(6) female patient who underwent mitral valve replacement with a 29 mm medtronic-hall mechanical valve.Although the valve disk moved freely, during closure a sudden drop in blood pressure occurred with electromechanical dissociation.The left atrium became noticeably distended.Cardiopulmonary bypass was reinitiated, and the atrium was re-opened.The tip of the posteromedial papillary muscle was wedged between the disk and valve ring, locking the valve in the closed position.The valve was opened manually, and the papillary muscle was debrided.The patient was weaned from cardiopulmonary bypass without difficulty and was discharged twelve days later.Patient 3: a (b)(6) female patient who underwent mitral valve replacement with a 31 mm medtronic-hall mechanical valve.The valve disk was seen to move freely, and the patient was weaned from cardiopulmonary bypass without difficulty.During closure of the sternotomy, recurrent episodes of electromechanical dissociation were noted.With a normal electrocardiogram, there were recurrent episodes of cessation of left ventricular ejection followed by spontaneous recovery.Cardiopulmonary bypass was reinitiated, and the atrium was opened.The tip of the posteromedial papillary muscle was found to lay on the valve ring, causing intermittent locking of the disk against the ring.The papillary muscle was debrided, resulting in a normally functioning valve.The patient was weaned from cardiopulmonary bypass and was subsequently discharged after an uneventful hospital course.No unique device identifier numbers were provided.No additional adverse patient effects or product performance issues were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDTRONIC HALL PROSTHETIC HEART VALVE
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
EI  
091708734
MDR Report Key13026563
MDR Text Key282404142
Report Number2025587-2021-03770
Device Sequence Number1
Product Code LWQ
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P790018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Literature,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberM7700
Device Catalogue NumberM7700
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/22/2021
Initial Date FDA Received12/17/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Life Threatening; Death;
Patient Age62 YR
Patient SexFemale
-
-