Model Number FLOW-I C20
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/08/2021
Event Type  malfunction  
Event Description
It was reported that the sevoflurane vaporizer had a yellow residue on the inside.The vaporizer is the unit containing anesthetic agent in the anesthesia workstation.There was no patient harm.Manufacturer´s ref.# (b)(4).
 
Event Description
Manufacturer's reference #: (b)(4).
 
Manufacturer Narrative
The investigation has been completed.The vaporizer was returned for investigation and the reported issue was confirmed.The yellow residue inside the vaporizer has been analyzed in a chemical analyze and the cause of the yellow residue is due to degradation of sevoflurane creating hydrogen fluoride.The concerned vaporizers are sevoflurane with maquet fill and safe-t-seal (which is keyed to sevoflurane from baxter).A root cause analysis has been performed regarding the formation of hydrogen fluoride within sevoflurane vaporizers.The root cause analysis found that changes in design and manufacturing process introduced in 2018 was the reason for the formation of hydrogen fluoride.The main reason for the manufacturing change in 2018 was to facilitate easier assembling in production and to get a better sourcing of certain components.The surface treatment in the aluminum canister that form the main part of the liquid container in the vaporizer was found to be inadequate.To prevent this from happening the sevoflurane vaporizer will be redesigned.The fsca has been initiated to remove all the concerned vaporizers from the field.The removing of the concerned vaporizers are to be completed until (b)(4) 2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOW-I
Type of Device
GAS-MACHINE, ANESTHESIA
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
sara soderholm roman
roentgenvagen 2
solna 
MDR Report Key13042865
MDR Text Key282489257
Report Number8010042-2021-03021
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeNO
PMA/PMN Number
K191027
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFLOW-I C20
Device Catalogue Number6677200
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/01/2021
Initial Date FDA Received12/20/2021
Supplement Dates Manufacturer Received07/14/2022
Supplement Dates FDA Received07/18/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Unknown
Removal/Correction NumberZ-0468-2022
Patient Sequence Number1
-
-