MEDTRONIC EUROPE SARL REVEAL LINQ; DETECTOR AND ALARM, ARRHYTHMIA Back to Search Results
Model Number LNQ11
Device Problem Inappropriate or Unexpected Reset (2959)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/29/2020
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.Performance data collected from the device was received and analyzed.Analysis of the device memory indicated a partial electrical reset.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that an internal review of data transmitted from the device indicated that a partial power on reset (por) occurred.The icm was explanted.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVEAL LINQ
Type of Device
DETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ   1131
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key13042913
MDR Text Key282492901
Report Number9614453-2021-04693
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00643169845749
UDI-Public00643169845749
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date 12/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/28/2020
Device Model NumberLNQ11
Device Catalogue NumberLNQ11
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2021
Initial Date Manufacturer Received 12/03/2021
Initial Date FDA Received12/20/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/30/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number2182208-052821-004-C
Patient Sequence Number1
Patient Age71 YR
Patient SexMale
-
-