ZIMMER BIOMET, INC. ARCOS 3.5MM HEX DRIVE; INSTRUMENT, HIP Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem Foreign Body In Patient (2687)
Event Date 11/26/2021
Event Type  Injury  
Manufacturer Narrative
(b)(4).Foreign: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the tip of arcos hex driver 3.5mm broke off during use.The fragment remains in the screw head.The fragment debris was firmly fixed to the screw head.The surgeon has decided that it couldn't be removed during the surgery and will be monitored for few weeks to few months.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4, b5, d4, d9, g3, g6, h2, h3, h4, h6, h10.Visual examination of the returned product identified the hex tip to be fractured.Wear rings are present around the shaft.The handle grip is scuffed and has a single discoloration spot.The connector is also scuffed.Additionally the features of the fracture surface visually align with a torsional overload fracture failure.Device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCOS 3.5MM HEX DRIVE
Type of Device
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13043451
MDR Text Key282502622
Report Number0001825034-2021-03388
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K090757
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number31-301852
Device Lot NumberZB161203
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/13/2021
Initial Date FDA Received12/20/2021
Supplement Dates Manufacturer Received01/31/2022
Supplement Dates FDA Received02/02/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/17/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-