Model Number BI70000028120 |
Device Problems
Failure to Charge (1085); Smoking (1585); Unexpected Shutdown (4019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/17/2021 |
Event Type
malfunction
|
Event Description
|
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that the mobile view station (mvs) dies immediately after being unplugged.The site contact checked on the unit this morning and when testing it by unplugging it from the wall they smelled a burning smell.Since then it has not lasted on battery power for more than a few seconds.
|
|
Manufacturer Narrative
|
Other relevant device(s) are: product id: b i71000480.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received additional information that the mobile view station (mvs) was left unplugged too long which caused the issue.After the site plugged the unit in fully, and left it overnight to charge, the system performed as intended.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|