Model Number OER-PRO
Device Problem Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/29/2021
Event Type  malfunction  
Event Description
Customer was having an issue with the device disinfectant (lcg cartridge) drawer closing.They were able to correct the problem, possible tray alignment, and finish the process with no error codes.There is no patient involvement and no harm reported to any patient.
 
Manufacturer Narrative
Field service engineer (fse) went on site to evaluate the customer reported issue of having trouble closing the lcg (disinfectant) drawer.Customer informed the fse that the drawer will not open or close correctly.Fse inquired if they were pulling off the stickers before loading the bottles in the drawer.Customer reported that it was being done.The drawer tracks and draw flow on the rails was were functioning normally.However, upon opening the drawer, fse observed that the stickers were still on the bottle.Fse explained to the customer proper process of loading bottles and how to close the drawer once by sliding up to the cutters before pushing draw closed with slight force to cut in to caps.They will follow the proper process of sticker removal and proper loading as well.No repair was required.Evaluation is ongoing.Supplemental report(s) will be submitted when any relevant new information is available.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely the disinfectant bottle drawer was not fully pushed in when replacing the disinfectant bottle.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13047641
MDR Text Key286538722
Report Number8010047-2021-16428
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/29/2021
Initial Date FDA Received12/20/2021
Supplement Dates Manufacturer Received02/22/2022
Supplement Dates FDA Received03/01/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/28/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-