MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPE REPROCESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-PRO
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2021
Event Type  malfunction  
Manufacturer Narrative
The field service engineer (fse) went on site to evaluate and repair the customer reported issue of the broken yellow connector.Fse found the yellow connector broken and replaced it.Equipment was repaired and verified according to original equipment manufacturer instructions by the fse.The covers of the oer-pro were not re moved so an electrical safety check was not required.Evaluation is ongoing.Supplemental report(s) will be submitted when any relevant new information is available.
 
Event Description
As reported by the customer, the yellow connector of the device was broken.There is no reported harm to any patient.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 7 years since the subject device was manufactured.Based on the results of the investigation, it is likely the yellow connector could not be connected as some impact was added to the connector and the connector became loose and came apart.Probable causes for the event include: -accumulated stress over time which caused the connector to get loose.-unintentional impact to the connector with something hard.The following information is stated in the instructions for use: ¿check the following for each connector.The connector should be fixed firmly.The o-rings should be free of abnormalities such as cracks, tears, or dents.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13048004
MDR Text Key286921263
Report Number8010047-2021-16433
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/30/2021
Initial Date FDA Received12/20/2021
Supplement Dates Manufacturer Received01/26/2022
Supplement Dates FDA Received02/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/09/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-