One device was returned for investigation.Under visual inspection, the sample appeared to be in good condition.An inflation test was done and after 12 hours of inflation, no leak was found.Device was also tested under water with no leak found as well.No lot number was provided so no dhr was done.
|
It was reported that the customer started to use the product from oct/12.Four (4) days later, a phenomenon that was suspected to be air leak from the cuff was observed.The product was checked, but since no anomaly in it was confirmed, use was continued until nov/06.Then, when the customer checked the product after the removal of it, leakage of air from the cuff was detected.No patient injury.No additional information is available for this complaint.
|